CO2 - Gas Sensor ModuleThe module CO2-4K-3M-P is designed to measure theconcentration of Carbon-dioxide in the air. This concentration is an reliable indicator for the “quality” of the air in e.g. a conference room. The higher the concentration of CO2 the worse is the quality of the air. High values of CO2 concentration in the air make a person tired and the mental performace will decrease. In “fresh air” the concentration of CO2 is 0.034% (vol) which corresponds to 340 ppm. The indoor concentration in offices should not exceed 1000 ppm, and the maximum allowable concentration is 5000 ppm. The CO2 module CO2-4K-3M-P is based on an optical mesurement principle. infrared radiation of wavelength 15 μm is emitted from a light source and is detected by a wavelength selective detector. Depending on the length of the optical path and the concentration of CO2 the light is absorbed (and re-emitted in all directions). Measurement range is 400 ppm ... 4000 ppm with a resolution of 20ppm. The response time is about 180 seconds and there is no pump necessary to drive the module. Output format is PWM. ![]() Interlocuteur
|